Product

DSP  AD5430-07(Thermocouple board)

Features

This board is for voltage inputs frorm thermocouple inputs (K, T, J). It provides 8 channels for thermocouple measurement in addition to the cold contact compensation. Measurement speed is 250 msec/8ch (4 measurements per second). The board consists of two sub-boards ( two slots required):,a main board and the a daughter board.

Specifications

Input channel number Thermocouple voltage input: 8ch
Measurement period Fixed at 250 msec/8ch
Measurement target Thermocouple: K, T, and J
Voltage: ±200 mV, ±2 V, ±20 V
Measurement resolution 0.1°
Reference contact compensation accuracy ±0.5° (when measuring 0° or higher)
Maximum permitted input voltage ±25 V DC
Insulation withstand voltage 1000 VAC (50/60 Hz) for 1 min.
Input resistance 10 MΩ or greater
Input external resistance 2 kΩ or smaller
Cold contact point compensation The input of the thermo sensor attached on the terminal board is read and used as the reference contact compensation for the thermocouple input.
Burnout detection Provides discontinuity detection for thermocouple sensors. When burned out, output swings fully to the positive side.

Measurement range

Range Measurement range Measurement accuracy
K -200.0 to 1370.0 Degree ±(0.15% of reading +0.7°)
For -200 to-100°, ±(0.15% of reading +1°)
T -200.0 to 800.0 Degree ±(0.15% of reading +0.5°)
For -200 to-100°, ±(0.15% of reading +0.7°)
J -200.0 to 1100.0 Degree
±200mV -200.0 to +200.0 mV ±(0.1% of reading +200µV)
±2V -2.000 to +2.000 V ±(0.1% of reading +2mV)
±20V -20.00 to +20.00 V ±(0.1% of reading +20mV)

Page Top