RA3100 Software Ver. 1.3.1 (April 2022) Magor Changes 1.Fixed a bug that waveform printing is faint. When the input frequency of external sampling is about 10Hz or less, the waveform printing will be faint in whole. 2.Fixed a bug that printing may not start even if waveform printing is started (PRINT key LED lights up). 3.Fixed a bug that caused a printer system error. A printer system error may occur during waveform printing and printing may stop. 4.Fixed a bug that the footer is not printed. If the waveform is printed with the channel assigned to the place where there is no division graph with the waveform division function, the footer will not be printed. Example: It occurs when the waveform division is set to 2 divisions and printed with channels assigned to Graph 10 (G10). 5.Fixed a bug that the PRINT key LED does not turn off. After stopping pen recording, waveform printing will stop, but the PRINT key LED may remain lit. 6.Fixed a bug that "0" of X axis is not printed. When pen-recording is started with the PRINT IN signal of the remote control module, "0" at the beginning of the X-axis of waveform printing may not be printed. 7.Fixed a bug that the print position of the trigger / mark shifts. If printing is started with the PRINT IN signal of the remote control module, the print position of the trigger / mark within 100 sample data from the start of printing may shift. 8.Fixed a bug that printing may stop if the sampling speed is changed during pen recording. 9.Fixed a bug that the PRINT key LED did not turn off and the "Saving" dialog did not close. When recording using real-time waveform printing, the PRINT key LED may not turn off even after recording is completed, and the "Saving" dialog may not close. 10.Fixed a bug that the sampling speed is not displayed in the control bar. When starting with a sampling speed of 10min / div, the sampling speed is not displayed in the control bar. 11.Fixed a bug that the selection item of the measurement range is not displayed. The selection items are not displayed even if the measurement range selection dialog of the [2CH temperature module] CH setting submenu is opened. 12.Fixed a bug that you can not update after 3 days have passed since the main unit started. 13.Fixed a bug that the soft keyboard disappears after 3 days have passed since the main unit started. For the update method, refer to the "RA3100 Update Procedure Manual".