Product

Weighing

  • Balances
  • Scales
  • Indicators
  • Loadcells
  • Peripherals
  • Software
LC-4103 (Aluminum)
For use with medium size platforms, single point compact load cells
LC-4103
Capacity Model No. CAD file
60 kgf LC4103-K060 CAD Download
100 kgf LC4103-K100
150 kgf LC4103-K150
Download 3D CAD
PDFSetup Manual (PDF 323KB)
Please send your inquiry here
Features
Parallelogram-type construction
400 × 600 mm platforms can be used
Temperature compensation maintains both zero and span from -10 °C to +40 °C (+14 °F to +104 °F)
Deadweight up to 80% of capacity. Load weight and deadweight up to 180% of capacity
Moisture proof
Compact and lightweight
Specifications
Rated capacities ..........................60 kgf, 100 kgf, 150 kgf
Rated output ..................................1 mV/V+15/-0%
Maximum deadweight .................80% of rated capacity
Maximum safe overload .............300% of rated capacity
Combined error ...........................±0.015% of rated output
Creep .............................................±0.05% of rated output/hour
Zero balance ...............................20 ± 5% of rated output
Compensated
temperature range .....................-10 °C to +40 °C/+14 °F to +104 °F
Recommended
excitation voltage .......................12 V
Maximum excitation
voltage .........................................15 V
input terminal resistance .........Approximately 400 Ω
Output terminal resistance .....350 Ω ± 5 Ω
Insulation resistance ...............Greater than 5000 MΩ at DC50 V
Temperature effect
Zero .............................................0.004% of rated output/°C
Span ............................................0.0014% of rated output/°C Typ.
Platform size .............................400 × 600 mm/15.8 × 23.6 inches