Item |
TM-2917 |
Blood pressure measurement |
Oscillometric method |
Operating Modes |
Set either Inflation or Deflation mode |
Patient Applications |
Adult, Children, Neonatal |
Pressure detection |
For measurement:
For safety circuit: |
Semiconductor pressure sensor
Semiconductor pressure sensor |
Measurement range |
Pressure:
Pulse Rate: |
0 to 299 mmHg
30 to 240 bpm |
Measurement Accuracy |
Pressure:
Pulse Rate: |
±3 mmHg over full range in operating conditions
±2% or ±2 BPM, whichever is greater |
Systolic Range |
(Inflation) Adult:
Neonatal:
(Deflation) Adult:
Neonatal: |
60 to 280 mmHg
N/A
40 to 280 mmHg
40 to 130 mmHg |
DIA Range |
(Inflation) Adult:
Neonatal:
(Deflation) Adult:
Neonatal: |
40 to 200 mmHg
N/A
20 to 200 mmHg
20 to 100 mmHg |
Scale interval |
Pressure:
Pulse Rate: |
1 mmHg
1 bpm |
Inflation method |
Pump |
Rolling pump
10 s or less/ 0→300 mmHg
|
Exhaust system |
Constant exhaust:
Rapid exhaust: |
Automatic deflation controlled by solenoid valve
Rapid exhaust by opening of solenoid valve for controlled exhaust and rapid exhaust |
Electrical rating |
Rated power supply voltage:
Power input: |
Direct current +5 to +12 V
Not more than 8 VA |
Serial communication |
Standardl:
Logic levels:
Data-flow control: |
Based on RS232C
TTL (3.3V) signal voltage level
RTS from NIBP MODULE only (Non-use CTS) |
Operating Conditions |
Temperature:
Humidity:
Air pressure: |
0˚C to 60˚C
Not more than 85%RH (no condensation)
700 - 1060hPa |
Storage environment |
Temperature:
Humidity:
Pressure: |
-20 to +70°C
95%RH or less (with no condensation)
700 - 1060hPa |
External dimensions |
60(H) x 39(W) x 104.5(D) |
Weight |
260g |
NIBP Accuracy |
Meets ISO 81060-2:2013 |
Safety & Regulatory Standards |
IEC 60601-1:2005+AMD1:2012, IEC 60601-1-2:2014, IEC 810601-2-30:2018 ISO 81060-2:2013, IEC 62304:2006 |
Service life |
Time
Cycles |
6 years
100,000
|